All PRODUCTS

TL-85EI

The autoclave is the equipment which utilizes the pressure saturated steam for sterilizing goods quickly and reliable. It is applicable to be used in medical and health work, scientific and agriculture institutes. It is a perfect equipment for sterilizing the medical equipment, dressings, glassware, liquor culture mediums, etc.

Description

Product Features:

PMMA I microcomputer intelligent operating system

Large-size capacitive touch screen enables real-time viewing of the operating parameters and working status diagram of the device during its working phase 

Positive pressure pulse, 6 level cold air discharge procedure can be set to ensure the thorough cold air discharge 

Equipped with 8 fixed programs +6 custom programs 

After sterilization, the 6-stage pressure relief rate is optional 

With three-level authority management of engineer, administrator and operator

The whole machine anti – hot design, heat insulation and beautiful 

Classic handwheel type door opening structure, with double protection of pressure mechanical self-locking and closed cover self-inspection 

Equipped with 100-hour countdown reservation function 

One key electric drainage, more convenient for operation 

Equipped with boiling point setting function to ensure that the instrument can be heated to the required sterilization temperature at high altitude 

Standard equipped with condenser + air cooling + secondary water cooling rapid cooling system to shorten the waiting time for opening the cover 

One waterproof socket, two 485 ports and G1/2A standard test pipe thread holes 

Fault detection function, overtemperature shutdown protection, water cut-off protection, overpressure self-relief protection, anti-hot protection, overload protection, 

Short circuit protection, leakage protection, independent electronic overvoltage protection and other multiple security

Optional accessories:

Special printer: realize date, time, temperature, sterilization process and curve printing 

Sample temperature probe: realize the printing of F0 value Verify the lead

|model:BXM – 85EI 

Sterilization Temp 105~135℃

Temp. fluctuations ˂±1℃

Temp. Uniformity ≤1℃

Melting Temp 50 ~ 100 ℃

Heat Preservation Temp40 ~ 60 ℃

Temperature resolution 0.1℃

Sterilization Timer Range 0~120min(0~6000min customized)

Melting Timer Range  0 ~ 6000 min

Heat Preservation Timer Range  0 ~ 6000 min

Reservation Range 0 ~ 6000 min

Design Pressure  0.38MPa

Working Pressure  0.22 MPa

Pressure gauge indicating range  0~0.4MPa (class 1.6)

Power (kW) 3.5

Power Supply ~220V±22V, 50HZ±1HZ

Environment Temp 5 ~ 40 ℃

Environment Humidity  20% ~ 80% RH

Housing Dimensions W×D×H (mm) 535×640×1165

Internal size (mm) Ø400×725

Use height (mm) 599

Volume: 85 l

Top Ø 360 * 270* 2

Gross weight /Net weight  (kg) 124/105

Package sizeW×D×H (mm)  780 x 680 x 1350